MAUDE Adverse Event Report: STRYKER MEDICAL-KALAMAZOO ISOTOUR, 84 IN,PREMIUM END406; MATTRESS, AIR FLOTATION, ALTERNATING PRESSURE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER MEDICAL-KALAMAZOO ISOTOUR, 84 IN,PREMIUM END406; MATTRESS, AIR FLOTATION, ALTERNATING PRESSURE Back to Search Results
Model Number 2872
Device Problems Patient-Device Incompatibility (2682); No Apparent Adverse Event (3189)
Patient Problems Pressure Sores (2326); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/09/2021
Event Type  malfunction  
Event Description
It was reported that while lying on the mattress, the patient experienced skin break down leading to a stage ii pressure injury.The user facility reported the injury was treated with pressure relieving interventions, moisture management, regular skin assessments, and daily mist therapy.Attempts have been made to gather additional information from the user facility but stryker has not received a response back at this time.
 
Manufacturer Narrative
The device evaluation has been completed and section h codes have been updated to reflect this.Additional information regarding the alleged injury has been provided by the user facility and updated under section b5.
 
Event Description
It was reported that a patient sustained a pressure injury while using the isotour surface in the icu.The user facility stated that all turning and injury prevention protocols were followed; however, due to the patient's skin condition and braden scores they were at risk for developing a pressure injury before being placed on the isotour surface.There are a number of different factors that contribute to pressure injuries, such as nutrition, skin condition, moisture and pressure injury treatment protocol.Given that all factors that are within stryker¿s control were evaluated and found to be operating as intended, it was determined that there were no defects found with the isotour surface that would have potentially caused or contributed to the alleged pressure injuries.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ISOTOUR, 84 IN,PREMIUM END406
Type of Device
MATTRESS, AIR FLOTATION, ALTERNATING PRESSURE
Manufacturer (Section D)
STRYKER MEDICAL-KALAMAZOO
3800 east centre avenue
portage MI 49002
MDR Report Key11961876
MDR Text Key255082089
Report Number0001831750-2021-00947
Device Sequence Number1
Product Code FNM
UDI-Device Identifier07613327424607
UDI-Public07613327424607
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 08/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/08/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number2872
Device Catalogue Number2872000011
Was Device Available for Evaluation? Yes
Date Manufacturer Received05/26/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-