• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. EVIS EXERA II ULTRASOUND GASTROVIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. EVIS EXERA II ULTRASOUND GASTROVIDEOSCOPE Back to Search Results
Model Number GF-UCT180
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The device was returned to an olympus service center for evaluation and the issue was confirmed.The insertion tube was sticky and there was a foreign black residue found in the forceps passage.In addition, the cover glue was peeling and there was a crack in the glue.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
 
Event Description
The user facility reported to olympus there was glue stuck on the insertion tube and inside the evis exera ii ultrasound gastrovideoscope.The issue was found at reprocessing.No patient involvement reported.
 
Manufacturer Narrative
This supplemental report is being submitted to provide the results of the legal manufacturer¿s investigation.The device history records for this device were reviewed and all records indicated that the product was manufactured according to all applicable procedures and met final product release criteria.No abnormalities were found.The root cause of the issue could not be conclusively specified.Judging from the manufacturing year of the subject device, it was unlikely that the peeling was due to aging deterioration.It was likely that the phenomenon occurred because external force was applied to the relevant part by strongly rubbing it during daily use including reprocessing.The instructions for use (ifu) provides the following guidelines: inspection of the endoscope 3.Inspect the external surface of the entire insertion section including the bending section and the distal end for dents, bulges, swelling, scratches, holes, sagging, transformation, bends, adhesion of foreign bodies, missing parts, forming objects, or other irregularities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II ULTRASOUND GASTROVIDEOSCOPE
Type of Device
ULTRASOUND GASTROVIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key11965124
MDR Text Key255120768
Report Number8010047-2021-07286
Device Sequence Number1
Product Code ODG
UDI-Device Identifier04953170341809
UDI-Public04953170341809
Combination Product (y/n)N
PMA/PMN Number
K093395
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 07/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberGF-UCT180
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/12/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/12/2021
Initial Date FDA Received06/09/2021
Supplement Dates Manufacturer Received07/05/2021
Supplement Dates FDA Received07/27/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-