• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. EVIS LUCERA ELITE GASTROINTESTINAL VIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. EVIS LUCERA ELITE GASTROINTESTINAL VIDEOSCOPE Back to Search Results
Model Number GIF-H290Z
Device Problem Use of Device Problem (1670)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/17/2021
Event Type  malfunction  
Manufacturer Narrative
Olympus medical systems corp.(omsc) could not investigate the subject device, because the subject device was not returned to omsc.Device history record review indicates that the product was manufactured and tested in accordance with all applicable procedures and met all final product release criteria.Since the subject device was not returned to omsc, the exact cause was unknown.Omsc surmised that the reported phenomenon occurred due to the following cause.-the user did not check that the reprocess was completed normally according to the instruction manual.
 
Event Description
Olympus medical systems corp.(omsc) was informed from the user that after the user used the subject device for the procedure, the subject device was not reprocessed and was used for next procedure.The subject device was placed in the reprocessing basin of the oer-3 but a staff did not start the reprocess program of the oer-3.Another stuff judged that the reprocess was completed and took it for the next procedure.Other detailed information was not provided.There was no report of patient injury associated with the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS LUCERA ELITE GASTROINTESTINAL VIDEOSCOPE
Type of Device
GASTROINTESTINAL VIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key11972255
MDR Text Key280677320
Report Number8010047-2021-07320
Device Sequence Number1
Product Code FDS
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial
Report Date 06/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/10/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberGIF-H290Z
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received05/18/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-