• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COCHLEAR LTD NUCLEUS 22; NUCLEUS 22 COCHLEAR IMPLANT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COCHLEAR LTD NUCLEUS 22; NUCLEUS 22 COCHLEAR IMPLANT SYSTEM Back to Search Results
Model Number CI22M
Device Problem Electrical Shorting (2926)
Patient Problem Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 05/20/2021
Event Type  Injury  
Manufacturer Narrative
This report is submitted on (b)(6) 2021.
 
Event Description
Per the clinic, the device was explanted on (b)(6) 2021, due to multiple short circuits.The patient was reimplanted with another cochlear device during the same surgery.
 
Manufacturer Narrative
Analysis revealed damaged wires along the helix section of the electrode lead.The device analysis report is attached.This report is submitted on september 15, 2021.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NUCLEUS 22
Type of Device
NUCLEUS 22 COCHLEAR IMPLANT SYSTEM
Manufacturer (Section D)
COCHLEAR LTD
1 university avenue
macquarie university, nsw 2109
AS  2109
MDR Report Key11981394
MDR Text Key255630579
Report Number6000034-2021-01750
Device Sequence Number1
Product Code MCM
Combination Product (y/n)N
PMA/PMN Number
P970051
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 08/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/11/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberCI22M
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/09/2021
Date Manufacturer Received08/25/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age31 YR
-
-