• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTECH SA ROSA ONE; COMPUTER-ASSISTED SURGICAL DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTECH SA ROSA ONE; COMPUTER-ASSISTED SURGICAL DEVICE Back to Search Results
Model Number OPTICAL DISTANCE SENSOR
Device Problem Device Damaged by Another Device (2915)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/20/2021
Event Type  malfunction  
Manufacturer Narrative
A full analysis of the data logs has been performed.This analysis concluded that the device shutdown was not provoked by the collision but is due to a communication error between the software and the controller.Indeed, the distance sensor cable was pinched in joint which caused an axis speed error and the controller shutdown procedure as expected.The design issue on distance sensor is known.The issues experienced will be addressed through the continuous improvement of our product.Unique identifier (udi) #: (b)(4).
 
Event Description
During registration, arm pinched distance sensor cable in joint 6, collision detected, registration needed to be restarted, delay 12 minutes, patient under anesthesia, no incision, seeg procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROSA ONE
Type of Device
COMPUTER-ASSISTED SURGICAL DEVICE
Manufacturer (Section D)
MEDTECH SA
zac eureka
900 rue du mas de verchant
montpellier, languedoc-roussillon 34000
FR  34000
Manufacturer (Section G)
MEDTECH SA
zac eureka
900 rue du mas de verchant
montpellier, languedoc-roussillon 34000
FR   34000
Manufacturer Contact
jay sharma
1520 tradeport drive
jacksonville, FL 32218
9047414400
MDR Report Key11981786
MDR Text Key255691175
Report Number3009185973-2021-00148
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K200511
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 06/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/11/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberOPTICAL DISTANCE SENSOR
Device Catalogue NumberROSAS00316
Device Lot NumberROSA3-277B
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received05/20/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/05/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
-
-