TERUMO CARDIOVASCULAR SYSTEMS CORPORATION BULK NON STER XCOAT DELPH PUMP; CENTRIFUGAL PUMP Back to Search Results
Model Number 3ZZ164275X
Device Problem Decoupling (1145)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/24/2021
Event Type  malfunction  
Manufacturer Narrative
Terumo has not received the device for evaluation; therefore, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.(b)(4).
 
Event Description
The user facility reported to terumo cardiovascular that prior to cardiopulmonary bypass, during prime, they noticed decoupling issue with delphin pump head.As per user it was "demagnetized" and wasn't functioning properly.No known impact or consequence to patient.Product was changed out.Procedure was completed successfully.
 
Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations.Upon further investigation of the reported event, the following information is new and/or changed: d4 (additional device information - added expiration date).G3 (date received by manufacturer).G6 (indication that this is a follow-up report).H2 (follow-up due to additional information and device evaluation).H3 ( device evaluated by manufacturer).H4 (device manufacture date).H6 (identification of evaluation codes 10, 11, 3331, 213, 67).Type of investigation #1: 10 - testing of actual/suspected device.Type of investigation #2: 11 - testing of device from same lot/batch retained by manufacturer.Type of investigation #3: 3331 - analysis of production records.Investigation findings: 213 - no device problem found.Investigation conclusions: 67 - no problem detected.The affected sample was inspected upon receipt with no visual anomalies noted.The affected sample and a representative retention sample from the same lot were setup to check flow rates at the outlet of the pump.The samples were tested at flow rates that span the pumps rated flow capability across the pumps rpm range.The pump outlet pressure was then measured at the respective flow and speed.The affected sample and retention sample performed as expected.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on june 11, 2021.Upon further investigation of the reported event, the following information is new and/or changed: h3 (device evaluation anticipated by manufacturer - a second follow-up will be submitted upon completion of the investigation and/or submission of new information, thus tcvs references conclusions code 11.) all available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BULK NON STER XCOAT DELPH PUMP
Type of Device
CENTRIFUGAL PUMP
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
125 blue ball road
elkton MD 21921
MDR Report Key11982820
MDR Text Key264321813
Report Number1124841-2021-00137
Device Sequence Number1
Product Code KFM
Combination Product (y/n)N
PMA/PMN Number
K112229
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Type of Report Initial,Followup,Followup
Report Date 08/06/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/11/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2022
Device Model Number3ZZ164275X
Device Catalogue NumberN/A
Device Lot NumberYA21
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/18/2021
Was the Report Sent to FDA? No
Date Manufacturer Received07/27/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-