• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ACTIS RETAINING STEM INSERTER; INSERTION DEVICES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ACTIS RETAINING STEM INSERTER; INSERTION DEVICES Back to Search Results
Model Number 2010-07-100
Device Problem Use of Device Problem (1670)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/01/2021
Event Type  malfunction  
Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that the threaded rod is missing on the device and should be code for missing components.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: examination of the returned instrument cannot confirm the complaint.Depuy-synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.H10 additional narrative:  added: d9 corrected: h3.
 
Manufacturer Narrative
Product complaint # (b)(4).All medical device reports associated with the same/similar device(s) and visual : missing components were reviewed.It was determined visual : missing components has not caused or contributed to any deaths or serious injuries within the time period of jan 1, 2018 ¿ july 12 2022.In total, there have been zero serious injuries and zero deaths reports related to visual : missing components in the last 4.5 years.Based on the data, the likelihood of a death or serious injury occurring as a result of the failure is remote; therefore, visual : missing components associated with same/similar device(s) of this report will no longer be reported as a medical device report (mdr) unless the event results in a reportable event per 21 cfr 803.50.D2b(procode).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIS RETAINING STEM INSERTER
Type of Device
INSERTION DEVICES
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kara ditty-bovard
700 orthopaedic dr.
warsaw, IN 46581-0988
6107428552
MDR Report Key11987261
MDR Text Key255810396
Report Number1818910-2021-12500
Device Sequence Number1
Product Code HWR
UDI-Device Identifier10603295393986
UDI-Public10603295393986
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 05/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/11/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2010-07-100
Device Catalogue Number201007100
Device Lot NumberABG53207
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/28/2021
Is the Reporter a Health Professional? No
Date Manufacturer Received06/16/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/24/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-