It was reported that, during procedure, the blade on the retrograde drill could not be deployed and the wire was broken.Bone hole was made according to standard procedure and all broken pieces were retrieved from the patient.The procedure was completed with a backup device and no significant delay.No further complications were reported.
|
Internal complaint reference: (b)(4).H3, h6: the device, used in treatment, was received for evaluation.There was no relationship found between the returned device and the reported incident.A visual inspection found that the suture passer tube was loaded with the correct monofilament, and the monofilament was not broken.There was no evidence of denting or deformation on the retrograde drill or guidewire.A functional evaluation found that the retrograde drill actuated appropriately.The guidewire was easily inserted into the retrograde drill and functioned as expected.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review concluded this was an isolated event.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A review of risk management files found that the reported failure was documented appropriately.The complaint was not verified and the root cause could not be determined since the reported malfunction could not be duplicated during the product evaluation process.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.
|