• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER HAND PIECE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER HAND PIECE Back to Search Results
Catalog Number ASKU
Device Problem Disconnection (1171)
Patient Problems Intraocular Pressure Decreased (4468); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 04/26/2021
Event Type  Injury  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A health professional reported that during a routine cataract extraction procedure the irrigation tubing disconnected from the phacoemulsification handpiece.The patient¿s eye pressure was affected.The surgeon stopped the procedure and realized that the handpiece had caused damage which required an additional vitrectomy procedure.The tubing was tested prior to the surgery and determined to be secure.The procedure was completed after a delay.Additional information was requested but not yet received.
 
Manufacturer Narrative
The handpiece was not returned for evaluation.The serial number (s/n) was not provided and could not be determined based on the information provided.Therefore, manufacturing information could not be obtained.The reported event could not be confirmed.The root cause of the reported event cannot be determined conclusively.The manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HAND PIECE
Type of Device
UNK
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
MDR Report Key11995594
MDR Text Key256162620
Report Number2028159-2021-00634
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
PMA/PMN Number
UNK
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 09/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/14/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberASKU
Device Lot NumberASKU
Was Device Available for Evaluation? No
Date Manufacturer Received08/20/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-