Catalog Number 295600-001 |
Device Problem
Material Integrity Problem (2978)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/26/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The freedom home ac power supply will be evaluated by syncardia.The results will be provided in a follow-up mdr.(b)(6).
|
|
Event Description
|
The freedom home ac power supply is a component that enables the freedom driver to be plugged into an external power source.The customer, a syncardia certified hospital, reported that the freedom home ac power supply had a damaged connector.
|
|
Manufacturer Narrative
|
Visual inspection confirmed the customer-reported issue as the connector outer housing was found to be cracked.It is likely that the damage to the connector outer housing was caused by rough handling; i.E., an excessive amount of force being applied to the connector and outer housing.Despite the physical damage, the unit passed functional testing.This issue will be monitored and trended as part of the customer experience process.Syncardia has completed its investigation and is closing this file.Ce 5545 follow-up report 1.
|
|