Received one single dpt kit.The reported event of leakage from snap-tab was confirmed.Leakage was observed from dpt snap tab area during leak test.Leakage occurred from a hole, approximately 0.10 in length, on the poppet.A crack of approximately 10mm in size was also observed at the drip chamber body, leakage occurred from the crack.No other visible damage was found from the returned unit.Although serious complications are relatively uncommon, the physician is advised to consider the potential benefits in relation to the possible complications.It is common clinical practice to inspect all products before use.These products are used by highly trained clinicians, experienced in identifying and mitigating any hazards that arise during use.In addition, these devices are typically used in intensive care units or operating rooms, where patients are closely monitored.Medication not being delivered consistently could cause a deterioration in the patients condition and may necessitate additional interventions.It is unknown whether user or procedural factors contributed to the stated event.In this case, there were no patient complications noted.There was no evidence to confirm that the product failure is related to a manufacturing or design defect.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis and any excursions above the control limits are assessed and documented as a part of the monthly review.
|
It was reported that, during use in the recovery room after an aortic valve replacement and coronary artery bypass, this disposable pressure transducer connected to a pump to administer drugs, leaked analgesic medication through the snap tab.The quantity of medication loss is unknown.The issue was solved replacing the device for another one.There was no allegation of patient injury.Patient demographics are not available.
|