• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. HANDPIECE, ELECTRIC DERMATOME

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. HANDPIECE, ELECTRIC DERMATOME Back to Search Results
Model Number N/A
Device Problem Output below Specifications (3004)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/19/2021
Event Type  malfunction  
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Telephone number: (b)(6).Review of the most recent repair record determined that the harness assembly plug was damaged and the motor had low speed.The motor was replaced and the plug was repaired and resolved the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing.Device is used for treatment.Review of complaint history identified additional similar complaints for the reported item and part and lot combination.Complaints are monitored through monthly complaint review in order to identify potential adverse trends.A definitive root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.The event is confirmed.
 
Event Description
It was reported that the device was in need of repair because the connector pin was bent.Occurred during kit inspection and there was no patient involvement.Evaluation of this device later revealed that the motor was running slow.No adverse events were reported as a result of this malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HANDPIECE, ELECTRIC DERMATOME
Type of Device
DERMATOME
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
christina arnt
56 e. bell drive
warsaw, IN 46582
5745273773
MDR Report Key12010713
MDR Text Key256551644
Report Number0001526350-2021-00705
Device Sequence Number1
Product Code GFD
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 06/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/16/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00882100100
Device Lot Number63037236
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/25/2021
Was the Report Sent to FDA? No
Date Manufacturer Received06/01/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/13/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
-
-