Model Number AIA-2000
Device Problems Fluid/Blood Leak (1250); Infusion or Flow Problem (2964)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/01/2021
Event Type  malfunction  
Manufacturer Narrative
Submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.Device evaluation by manufacturer: a field service engineering (fse) was at the customer's site to address reported event.Fse confirmed reported error by reviewing the error log and reproduced the error by performing a diluent prime.Fse found diluent check valve leaking and replaced it.Fse successfully validated instrument by performing quality control run without error and within acceptable range.No further action required by field service.The aia-2000 instrument is functioning as expected.A 13-month complaint history review and service history review through aware date of event for similar complaints was performed for serial number (b)(4).There were two (2) similar complaints identified during the searched period, which includes this event.The aia-2000 operator's manual under appendix 4: error messages states the following: [2085] air detected during diluent suction by main arm cause: during diluent suction, it was determined that the tip failed to touch the liquid surface.If retry fails, the measurement result will be flagged (ds flag).Solution: contact tosoh service center or local representatives.The most probable cause of the reported event was due to failure of the diluent check valve.
 
Event Description
A customer reported getting error message "2085 air detected during diluent suction by main arm and leaking fluid" on the aia-2000 instrument.Instrument is down.A field service engineer (fse) was dispatched to address the reported event, which resulted in a delayed reporting of patient samples for estradiol (e2), luteinizing hormone (lhii), follicle stimulating hormone (fsh), beta human chorionic gonadotropin (bhcg) and prolactin (prl).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIA-2000
Type of Device
AIA-2000
Manufacturer (Section D)
TOSOH CORPORATION
shiba-koen first building
3-8-2 shiba
minato-ku, 10586 23
JA  1058623
Manufacturer (Section G)
TOSOH CORPORATION
shiba-koen first building
3-8-2 shiba
minato-ku, 10586 23
JA   1058623
Manufacturer Contact
bernadette oconnell
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo, japan 10586-23
JA   1058623
MDR Report Key12010846
MDR Text Key256734289
Report Number8031673-2021-00187
Device Sequence Number1
Product Code KHO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K971103
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 06/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/16/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAIA-2000
Device Catalogue Number022100
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/01/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/01/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse