Submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.Device evaluation by manufacturer: a field service engineering (fse) was at the customer's site to address reported event.Fse confirmed the complaint by looking at the computer screen.Fse found the hard drive had failed.Fse resolved the complaint by replacing the computer and monitor.Fse successfully validated the instrument by performing quality control run, which completed without error and within acceptable range.No further action required by field service.The aia-2000 instrument is functioning as expected.A 13-month complaint history review and service history review through aware date of event for similar complaints was performed for serial number (b)(4).There were no similar complaints identified during the search period.The most probable cause of the reported event was due to failure of the computer hard drive.
|