• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TOSOH CORPORATION AIA-2000

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TOSOH CORPORATION AIA-2000 Back to Search Results
Model Number AIA-2000
Device Problem Computer Operating System Problem (2898)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/21/2021
Event Type  malfunction  
Manufacturer Narrative
Submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.Device evaluation by manufacturer: a field service engineering (fse) was at the customer's site to address reported event.Fse confirmed the complaint by looking at the computer screen.Fse found the hard drive had failed.Fse resolved the complaint by replacing the computer and monitor.Fse successfully validated the instrument by performing quality control run, which completed without error and within acceptable range.No further action required by field service.The aia-2000 instrument is functioning as expected.A 13-month complaint history review and service history review through aware date of event for similar complaints was performed for serial number (b)(4).There were no similar complaints identified during the search period.The most probable cause of the reported event was due to failure of the computer hard drive.
 
Event Description
A customer reported that the aia-2000 computer screen was black.The customer verified all cables were plugged in correctly.Technical support specialist (tss) instructed the customer to reboot but that did not resolve the problem.A field service engineer (fse) was dispatched to address the reported event, which resulted in a delayed reporting of patient samples for intact parathyroid hormone (ipth).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIA-2000
Type of Device
AIA-2000
Manufacturer (Section D)
TOSOH CORPORATION
shiba-koen first building
3-8-2 shiba
minato-ku, 10586 23
JA  1058623
Manufacturer (Section G)
TOSOH CORPORATION
shiba-koen first building
3-8-2 shiba
minato-ku, 10586 23
JA   1058623
Manufacturer Contact
bernadette oconnell
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo, japan 10586-23
JA   1058623
MDR Report Key12010881
MDR Text Key256561036
Report Number8031673-2021-00179
Device Sequence Number1
Product Code KHO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K971103
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 06/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/16/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAIA-2000
Device Catalogue Number022100
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/01/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/21/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/01/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-