Model Number 2581-11-000 |
Device Problems
Material Twisted/Bent (2981); Scratched Material (3020)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/04/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Product complaint #: (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
It was reported that the instruments were old and worn.
|
|
Event Description
|
Additional information received indicated that there was no patient issues and the instruments are old and have scratches on them.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthese joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthese joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Manufacturer Narrative
|
Product complaint # : (b)(4).Investigation summary : the device associated with this report was not returned, thus the reported event could not be confirmed.The investigation could not verify or identify any product contribution to the reported event with the information provided.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
|
Search Alerts/Recalls
|