It was reported that, during an arthroscopic procedure, when the "8.0mm acufex trunav retrograde drill" was activated to open the blade, it was found that the material was already broken and the cutter blade could not be opened.There was no break inside the patient.The procedure was successfully completed without delay using a back-up device.No further complications were reported.
|
H3, h6: the reported device, used in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review concluded this was an isolated event.A relationship, if any, between the subject device and the reported event could not be determined.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
|