MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION CAROTID WALLSTENT; STENT, CAROTID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION CAROTID WALLSTENT; STENT, CAROTID Back to Search Results
Model Number 26606
Device Problem Device-Device Incompatibility (2919)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/26/2021
Event Type  malfunction  
Manufacturer Narrative
Age at time of event: 18 years or older.(b)(6).
 
Event Description
It was reported that the device became caught in the stent strut.The 70% stenosed target lesion was located in the mildly tortuous carotid artery.A 190cm filterwire ez and a carotid wallstent self expanding stent were selected for use.During the removal, the physician felt resistance at hand.The physician thought the filter part was caught in the strut at the tip of the stent.Resistance at hand disappeared when it was pulled as it was and was removed from the body.Subsequently, the device was found to be torn when taken out of the retrieval sheath.There were no issues with the usage or fluoroscopy during the procedure and no patient complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAROTID WALLSTENT
Type of Device
STENT, CAROTID
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC IRELAND LIMITED
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key12033445
MDR Text Key257196384
Report Number2134265-2021-07858
Device Sequence Number1
Product Code NIM
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial
Report Date 06/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/21/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number26606
Device Catalogue Number26606
Device Lot Number0026991387
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/26/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/18/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-