TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS- ADVANCED PERFUSION SYSTEM 1
|
Back to Search Results |
|
Model Number 893137 |
Device Problems
Inaccurate Flow Rate (1249); Output Problem (3005)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/27/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The service repair technician (srt) reported that during testing of the device at the service center, the flowmeter set point with the flow motor fully clock wise was 10.The specification for the flowmeter high limit is five, therefore the flowmeter flow was not within specification.During laboratory analysis, the product surveillance technician (pst) installed the flowmeter into a lab use only (luo) epgs.Verification/release testing was conducted and the pst observed the flowmeter to function as it should.
|
|
Event Description
|
Upon receipt of the device, the user reported that the electronic patient gas system (epgs) flowmeter failed the flow set point testing.There was no patient involvement.
|
|
Manufacturer Narrative
|
The reported complaint was confirmed.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|