Other text: investigation completed on a smiths pressure monitoring|medex transtar kits.The complaint of pressure sensor found to be broken was not confirmed.Five pictures were reviewed and shoed female luers disconnected from tube and a pressure transducer broken (transtar and trigger flush), however these parts were not sent for analysis.Visual inspection could not confirm complain on sample provided, as no damage was observed.Assembly revealed product passed 100 percent prior to release.Root cause: the root cause of the reported issue cannot be determined, as the returned sample did not exhibited breakage and the samples from pictures were not provided for evaluation in order to identify the cause of the damage.
|