• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. 12.5MM ENTRY REAMER; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. 12.5MM ENTRY REAMER; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Model Number 71631116
Device Problems Break (1069); Dull, Blunt (2407)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/04/2021
Event Type  malfunction  
Event Description
It was reported that a 12.5mm entry reamer was chipped and dull.As this was noticed upon field inspection, there was not patient involvement.
 
Manufacturer Narrative
The device, intended for use in treatment, was returned for evaluation.A visual inspection of the returned device confirmed the stated failure mode.The device shows signs of wear and tear and the tip is broken, rendering the device inoperable.The device shows signs of extensive use.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal any manufacturing or material abnormalities that could have caused or contributed to the reported incident.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however, we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
12.5MM ENTRY REAMER
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key12042945
MDR Text Key257520677
Report Number1020279-2021-05318
Device Sequence Number1
Product Code JDI
UDI-Device Identifier03596010419583
UDI-Public03596010419583
Combination Product (y/n)N
PMA/PMN Number
K123598
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 06/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/22/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71631116
Device Catalogue Number71631116
Device Lot Number13HM18077
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/14/2021
Date Manufacturer Received06/28/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-