Model Number 37612 |
Device Problems
High impedance (1291); Therapy Delivered to Incorrect Body Area (1508); Low impedance (2285)
|
Patient Problems
Paresthesia (4421); Unspecified Nervous System Problem (4426); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/25/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the patient used to just top off every couple of days.He was now charging every day for at least an hour (1.7h, 5 h, 2.7hr) to get to full each day.No known falls.No impedance issues seen at the appointment at the time.Historical impedances are similar.Patient has 2 groups.Group a and b use left 1+2-3- and right 2+3- but group a pw is 210 and group b is 190.Patient had changed from group b to a recently because he thought he was to be on it and was noticing some dystonia.One of the groups provides better speech results.Caller will retake impedances with pressure/position changes.Caller requested patient to try group b again to see about charging pattern.Additional information was received that patient was experiencing drain on the rechargeable battery that is not typical.Cause is unknown at time of report.Battery was replaced and was being sent for analysis by medtronic.Patient is also experiencing tingling sensation on right side.Left lead shows possible short circuit between contact 1and 2 at 192 ohms.Right side shows high impedance on contact 8 at 3700 ohms.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
H3.Analysis of the ins (s/n (b)(6) found no significant anomaly.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|