SMITH & NEPHEW, INC. NS VIS ADPT GUIDE LGNP KIT; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
Model Number V0200109 |
Device Problems
Inadequacy of Device Shape and/or Size (1583); Fitting Problem (2183)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/10/2021 |
Event Type
malfunction
|
Event Description
|
It was reported, that during a tka surgery, a ns vis adpt guide lgnp kit did not work properly; the tibia fit was insufficient.Surgery was completed, without significant delay, by using a back-up device.The patient was not harmed beyond what has been described.
|
|
Manufacturer Narrative
|
H3, h6: the associated device, used in treatment, was not returned for evaluation but the reported event could be confirmed through the engineering evaluation.There was over segmentation in multiple areas of the tibia during the design process.The engineer will be retrained by the segmentation subject matter experts to procedure (b)(4): visionaire tka segmentation standards.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.The potential probable cause for this event is a design error.Based on this investigation, the need for corrective action is indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.
|
|
Search Alerts/Recalls
|
|
|