Model Number CI-1500-01 |
Device Problems
Electrical /Electronic Property Problem (1198); Therapeutic or Diagnostic Output Failure (3023); Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
Patient Problem
Failure of Implant (1924)
|
Event Date 05/29/2021 |
Event Type
malfunction
|
Event Description
|
The recipient is reportedly experiencing loss of lock.External equipment was exchanged, however, the issue did not resolve.A ct scan revealed no abnormality.Revision surgery will be scheduled.
|
|
Manufacturer Narrative
|
The recipient's device was explanted.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Correction: the recipient was reportedly reimplanted with another advanced bionic cochlear device.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed the electrode was severed.This is believed to have occurred during revision surgery.In addition, this inspection revealed a dented bottom cover.The photographic imaging inspection revealed loose components and disturbed bond wires.System lock could not be obtained at any spacing.The electrode condition prevented an electrical test from being performed.The device passed the electrical tests performed.The device passed the mechanical test performed.The reported complaint of loss of lock with the device was verified during this analysis.The device was received with multiple cracks along the hybrid, disturbed wire bonds, and dislodged electrical components.A corrective action was implemented.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|