• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ADVANCED BIONICS, LLC HIRES 90K¿ ADVANTAGE IMPLANT; COCHLEAR IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ADVANCED BIONICS, LLC HIRES 90K¿ ADVANTAGE IMPLANT; COCHLEAR IMPLANT Back to Search Results
Model Number CI-1500-01
Device Problems Electrical /Electronic Property Problem (1198); Therapeutic or Diagnostic Output Failure (3023); Loosening of Implant Not Related to Bone-Ingrowth (4002)
Patient Problem Failure of Implant (1924)
Event Date 05/29/2021
Event Type  malfunction  
Event Description
The recipient is reportedly experiencing loss of lock.External equipment was exchanged, however, the issue did not resolve.A ct scan revealed no abnormality.Revision surgery will be scheduled.
 
Manufacturer Narrative
The recipient's device was explanted.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Manufacturer Narrative
Correction: the recipient was reportedly reimplanted with another advanced bionic cochlear device.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Manufacturer Narrative
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Manufacturer Narrative
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed the electrode was severed.This is believed to have occurred during revision surgery.In addition, this inspection revealed a dented bottom cover.The photographic imaging inspection revealed loose components and disturbed bond wires.System lock could not be obtained at any spacing.The electrode condition prevented an electrical test from being performed.The device passed the electrical tests performed.The device passed the mechanical test performed.The reported complaint of loss of lock with the device was verified during this analysis.The device was received with multiple cracks along the hybrid, disturbed wire bonds, and dislodged electrical components.A corrective action was implemented.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIRES 90K¿ ADVANTAGE IMPLANT
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
Manufacturer Contact
lacey o'linn
28515 westinghouse place
valencia, CA 91355
MDR Report Key12081970
MDR Text Key261693078
Report Number3006556115-2021-00972
Device Sequence Number1
Product Code MCM
UDI-Device Identifier07630016815805
UDI-Public(01)07630016815805(11)170913(17)190831
Combination Product (y/n)N
Reporter Country CodeIR
PMA/PMN Number
P960058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 05/31/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/29/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date08/31/2019
Device Model NumberCI-1500-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/13/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age22 YR
Patient SexMale
-
-