Updated fields: d4, d9, g3, g6, h2, h3, h6, h10.The mayfield modified skull clamp (a1059) was returned for evaluation.Failure analysis - the device was not fit for clinical use as the rollover to lock is binding and may not have been correctly locked when in use.The device required general service, replacement of damaged torque screw, and ratchet extension which are >10 years old and other parts.Root cause - the reported complaint was confirmed.Evaluation showed that parts of the device are damaged and worn.The definite root cause of the observed condition cannot be reliably determined but it is likely caused by wear and tear or improper handling.The unit needs repair, and replacement of worn parts.General maintenance and cleaning are required.Device exceeds its expected life of 7 years (sold on feb 25, 2020) and replacement is highly recommended.
|