• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. EVOS 4.7MM X 40MM OST SCR PT; PLATE, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. EVOS 4.7MM X 40MM OST SCR PT; PLATE, FIXATION, BONE Back to Search Results
Model Number 72434740
Device Problem Difficult to Insert (1316)
Patient Problems Bone Fracture(s) (1870); Bone Shedding Debris (4557)
Event Date 06/16/2021
Event Type  Injury  
Event Description
It was reported that, during an internal fixation, a evos 4.7mm x 40mm osteopenia screw partially threaded (case-(b)(4)) and a evos 4.7mm x 42mm osteopenia screw partially threaded (case-(b)(4)) were not suitable to fix all medial malleolus fractures and destroyed the smaller fragments of the bone.He states that the thread is difficult to start because it is not self tapping, and that he preferred the old sized 4.0mm cancellous screws from our previous orif set tc-100.Surgery was resumed, after a non-significant delay, with a back-up device.Current state of the patient is unknown.
 
Manufacturer Narrative
H3, h6: the device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.The clinical / medical investigation concluded that, without the requested clinical information or the device, the root cause of the insertion difficulty cannot be determined.Per the surgeon, ¿the thread was difficult to start because it was not self-tapping, in addition, the surgeon preferred the old sized 4.0mm cancellous screws from our previous orif set tc-100.¿ subsequently, the surgery was resumed, after a non-significant delay, with a back-up device.However, the patient¿s status is unknown.Therefore, the impact to the patient beyond that which has already been reported cannot be determined.Should any additional medical information be provided, this complaint will be re-assessed.A complaint history review found related failures for the listed device; this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file and instructions for use documents revealed this failure mode was previously identified.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Some potential probable causes for this event could include but not limited to user error or procedural variance.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVOS 4.7MM X 40MM OST SCR PT
Type of Device
PLATE, FIXATION, BONE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key12101214
MDR Text Key259433345
Report Number1020279-2021-05598
Device Sequence Number1
Product Code HRS
UDI-Device Identifier00885556637722
UDI-Public00885556637722
Combination Product (y/n)N
PMA/PMN Number
K172262
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 08/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/01/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number72434740
Device Catalogue Number72434740
Was Device Available for Evaluation? No
Date Manufacturer Received08/04/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-