It was reported that stylus of the programmer was unable to be used.The stylus was replaced, however the issue persisted.The programmer displayed a black screen with a "calibrate stylus" message.It was further reported that the programmer displayed a "stylus is out of range" message, and that the programmer was in bad condition the programmer was unable to boot up.It was also reported that the issue did not resolve after calibrating the stylus.The programmer was returned for service.There was no patient involvement.
|
Product analysis :analysis was able to confirm the customer comment that the stylus of the programmer was unable to be used, and that the programmer was unable to be calibrated.It was noted that the keyboard and the keyboard rails were broken.It was further noted that the lower case was worn out and the upper and lower display bullets were missing.Reconfigured hard drive and reloaded software.All found defective parts were replaced and all other identified issues were resolved.The programmer then passed final functional and system tests.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|