Model Number 461055E
Device Problems Break (1069); Disconnection (1171); Leak/Splash (1354); Material Too Rigid or Stiff (1544); Detachment of Device or Device Component (2907); Infusion or Flow Problem (2964); Pressure Problem (3012); Material Split, Cut or Torn (4008)
Patient Problem Insufficient Information (4580)
Event Date 01/28/2021
Event Type  malfunction  
Event Description
Reporting multiple product concern reports: enterel feeding tube with iris technology 10 french - 55 inches: the tubing has a port that can become disconnected when in use.There is no way to secure the port (it is under a luer lock), thus tube feeding solution can end up not infusing into the patient.Once this port becomes disconnected, the pump will continue to pump formula all over the patient and the floor.Nurses have tried to tape the port shut to keep it from spilling but it does not secure the tubing correctly.When attempting to change the tubing, the tube feed seemed to have hardened around the port.When attempting to unscrew the tubing, the rubber portion of the dobhoff ripped, and the top plastic part came off with the tubing.Was unable to continue tube feeds, and the port was leaky now.Enteral feeding tube with iris technology 10 french 55 inches: one of the port access became broken when too much pressure was exerted from running tube feed.Tube feeding was attached to access labeled "enteral fluids only".End of broken access port taped and tube feed restarted without further issues.10 french 43 inches: one of the two luer lock port broke, causes leaking and issues with ng tube.Other valve not broken but unable to give feedings with other broken port due to loss in pressure.When attaching tube feeding to port you cannot tighten to tight if not it can cause this issue of the port breaking.10 french 55 inches: port for enteral tube feeding broke during tube feed administration.12 french 43 inches: the purple port tore when rn was trying to pull cap off the medication port to administer meds.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KANGAROO
Type of Device
TUBES, GASTROINTESTINAL (AND ACCESSORIES)
Manufacturer (Section D)
CARDINAL HEALTH, INC.
15 hampshire street
mansfield MA 02048
MDR Report Key12108473
MDR Text Key259676400
Report Number12108473
Device Sequence Number1
Product Code KNT
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 06/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/02/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number461055E
Device Catalogue Number461055E
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA06/08/2021
Event Location Hospital
Date Report to Manufacturer07/02/2021
Type of Device Usage Unknown
Patient Sequence Number1
-
-