MEDIVATORS DSD EDGE; AUTOMATED ENDOSCOPE REPROCESSOR Back to Search Results
Model Number 78400-409
Device Problem Smoking (1585)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/02/2021
Event Type  malfunction  
Manufacturer Narrative
A medivators field service engineer arrived onsite to inspect the dsd edge aer and found several internal components were damaged.The damage observed to the unit is indicative of a short circuit.The user facility elected to permanently remove the unit from service in lieu of repairing.Medivators requested the unit to be returned for evaluation; however, the user facility declined.A replacement dsd edge aer was installed at the user facility.A 1-year complaint review indicates this to be an isolated event.No additional issues have been reported.
 
Event Description
The user facility reported that their dsd edge automated endoscope reprocessor (aer) began emitting smoke from the compressor during a cycle.No report of injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DSD EDGE
Type of Device
AUTOMATED ENDOSCOPE REPROCESSOR
Manufacturer (Section D)
MEDIVATORS
14605 28th ave n
plymouth MN 55447
Manufacturer (Section G)
MEDIVATORS
14605 28th ave n
plymouth MN 55447
Manufacturer Contact
daniel davy
14605 28th ave n
plymouth, MN 55447
4403927453
MDR Report Key12110220
MDR Text Key259721590
Report Number2150060-2021-00010
Device Sequence Number1
Product Code FEB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K092387
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 07/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/02/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number78400-409
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/02/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-