SMITH & NEPHEW, INC. FEMORAL HEAD/NECK IMPACTOR; PROSTH, HIP, SEMI-CONSTRA, UNCEMENT, METAL/POLY, NON-POROUS, CALICUM-PHOSPHATE
|
Back to Search Results |
|
Model Number 71360093 |
Device Problem
Break (1069)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/18/2021 |
Event Type
malfunction
|
Event Description
|
It was reported that, during set up, before an unknown treatment, the tip of one (1) femoral head/neck impactor was broken.This happened before use in patient.Treatment was performed, without any delay, using a back-up device instead.No health consequences reported.
|
|
Manufacturer Narrative
|
H6: the device, intended for use in treatment, was returned for evaluation.A visual inspection of the returned device confirmed the stated failure mode.The device is fractured and heavily damaged, rendering it inoperable.The fractured pieces were not returned.A review of the device history records for the listed batch did not reveal any deviation from the standard manufacturing processes.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.The device is a reusable instrument that can be exposed to numerous surgeries and cleaning cycles.As plastics are vulnerable and crack may have initiated during use and possible causes could be due to the heating and cooling associated with autoclaving or prolonged use.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however, we will continue to monitor future complaints and investigate, as necessary.We consider this investigation closed.
|
|
Search Alerts/Recalls
|
|
|