Model Number 71442145
Device Problem Corroded (1131)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/22/2021
Event Type  malfunction  
Event Description
It was reported that, during the set up of a tka surgery, the top ring of a knee balancer torque indicator was rusted.Surgery was performed, without any delay, with a back-up device instead.Patient was not injured as consequence of this problem.
 
Manufacturer Narrative
H6: the device, used in treatment, was returned for evaluation.A visual inspection of the returned device confirmed the stated failure mode.The device was found to have discoloration, rendering the device inoperable.The device shows signs of extensive use.A review of complaint history did not reveal additional complaints.A review of the manufacturing records did not reveal any manufacturing or material abnormalities that could have caused or contributed to the reported incident.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KNEE BALANCER TORQUE INDICATOR
Type of Device
PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key12125631
MDR Text Key260226781
Report Number1020279-2021-05764
Device Sequence Number1
Product Code HRY
UDI-Device Identifier03596010566775
UDI-Public03596010566775
Combination Product (y/n)N
PMA/PMN Number
K121393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 08/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/06/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71442145
Device Catalogue Number71442145
Device Lot Number09HM00475
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/09/2021
Date Manufacturer Received07/30/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-