Model Number 31535 |
Device Problem
Material Separation (1562)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/10/2021 |
Event Type
malfunction
|
Event Description
|
It was reported that the wire was fractured.A flexima drainage catheter was selected for use on bile duct blockage.During the procedure, after the device was delivered, the guidewire was fractured in the patient.The device was simply pulled out and completely removed from the patient's body and the procedure was completed with another of the same device.No complications were reported and the patient was stable post procedure.
|
|
Manufacturer Narrative
|
Device evaluated by mfr: the device was returned for analysis.The catheter was returned with the several accessories; moreover, the 0018" stainless guidewire was observed kinked and broken at the distal end section.This guidewire was received without the hoop protector.The other devices and accessories were not presented with any issue or damage.Dimensional inspection of the guidewire was performed and the measured dimensions were within specification.Under microscope it was possible to observe that the stainless guidewire was kinked and broken at the distal end section.Media analysis was performed based on a photo attached to the complaint record and the stainless guidewire was observed kinked and broken at the distal end section.No other issues were identified during the product analysis.
|
|
Event Description
|
It was reported that the wire was fractured.A flexima drainage catheter kit was selected for use on bile duct blockage.During the procedure, after the device was delivered, the guidewire was fractured in the patient.The device was simply pulled out and completely removed from the patient's body and the procedure was completed with another of the same device.No complications were reported and the patient was stable post procedure.
|
|
Search Alerts/Recalls
|