MAUDE Adverse Event Report: SMITH & NEPHEW, INC. GII QUICK CONNECT HANDLE; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. GII QUICK CONNECT HANDLE; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Model Number 71440044
Device Problem Loose or Intermittent Connection (1371)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/28/2021
Event Type  malfunction  
Event Description
It was reported that, during a tka a gii quick connect handle did not lock onto tibial tray anymore.Surgery was resumed, without any delay, with a smith and nephew back-up device.Patient was not harmed as consequence of this problem.
 
Manufacturer Narrative
The device, used in treatment, was returned for evaluation.A visual inspection of the returned device confirmed the stated failure mode.The device locking mechanism is worn, rendering the device inoperable.The device shows signs of extensive use.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal any manufacturing or material abnormalities that could have caused or contributed to the reported incident.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII QUICK CONNECT HANDLE
Type of Device
PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key12135588
MDR Text Key260484482
Report Number1020279-2021-05824
Device Sequence Number1
Product Code HRY
UDI-Device Identifier03596010213860
UDI-Public03596010213860
Combination Product (y/n)N
PMA/PMN Number
K121393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 08/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/08/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/21/2014
Device Model Number71440044
Device Catalogue Number71440044
Device Lot Number04FM05880
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/19/2021
Date Manufacturer Received08/26/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-