COCHLEAR LIMITED NUCLEUS CI612 COCHLEAR IMPLANT WITH CONTOUR ADVANCE ELECTRODE; NUCLEUS 24 COCHLEAR IMPLANT SYSTEM Back to Search Results
Model Number CI612
Device Problem Electrical Shorting (2926)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/09/2021
Event Type  Injury  
Manufacturer Narrative
This report is submitted on july 12, 2021.
 
Event Description
Per the clinic the device was explanted on (b)(6) 2021, due to loss of electrode function, intermittent function and short circuits on the electrode.The patient was reimplanted with another cochlear device during the same surgery.
 
Manufacturer Narrative
Device analysis report (dar) is attached this report is submitted on sep 07, 2021.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NUCLEUS CI612 COCHLEAR IMPLANT WITH CONTOUR ADVANCE ELECTRODE
Type of Device
NUCLEUS 24 COCHLEAR IMPLANT SYSTEM
Manufacturer (Section D)
COCHLEAR LIMITED
1 university avenue
macquarie university
nsw 2109
AS  2109
MDR Report Key12148628
MDR Text Key260933755
Report Number6000034-2021-02057
Device Sequence Number1
Product Code MCM
UDI-Device Identifier09321502036818
UDI-Public(01)09321502036818(11)210417(17)230416
Combination Product (y/n)N
PMA/PMN Number
P970051
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 08/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/12/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date04/16/2023
Device Model NumberCI612
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/23/2021
Date Manufacturer Received08/18/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age10 YR
-
-