• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AESCULAP AG CHALLENGER TI-P ML-LIGAT.CLIPS 12 CARTR.; LIGATION / VESSEL CLIPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AESCULAP AG CHALLENGER TI-P ML-LIGAT.CLIPS 12 CARTR.; LIGATION / VESSEL CLIPS Back to Search Results
Model Number PL579T
Device Problem Material Rupture (1546)
Patient Problem Insufficient Information (4580)
Event Type  malfunction  
Manufacturer Narrative
Manufacturing site evaluation: investigation on-going.Should relevant additional information / investigation results become available, a supplemental medwatch report will be submitted.
 
Event Description
It was reported to aesculap ag that a challenger ti-p ml-ligat.Clips 12 cartr (part # pl579t) was used during a procedure performed on an unknown date.According to the complainant, the clip pin broke and detached into the patient.The part was not able to be recovered.The fragment remained in situ.Additional information has been requested but not yet received as of this report.Additional patient information is not available.The malfunction is filed under aag reference (b)(4).
 
Manufacturer Narrative
Investigation results: as of the date of this report the complaint product was not provided for investigation.Therefore, a thorough investigation is not possible.Batch history review: the device quality and manufacturing history records have been checked for the available lot number and were found to be according to our specifications valid at the time of production.Review of the complaint history revealed that no similar complaints have been filed against products from this batch number.The review of risk assessment revealed that the overall risk level (severity 4(5) x probability of occurrence 1(5)) according to din en iso 14971 is still acceptable.Conclusion and measures / preventive measures: based upon the investigation results a clear root cause conclusion cannot be drawn.There is no indication for a material-, manufacturing- or design-related failure.In the event that the complaint product will be provided for investigation in the future, an update of this report will be provided unsolicited.Based upon the investigations results a capa is not necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CHALLENGER TI-P ML-LIGAT.CLIPS 12 CARTR.
Type of Device
LIGATION / VESSEL CLIPS
Manufacturer (Section D)
AESCULAP AG
po box 40
tuttlingen, 78501
GM  78501
MDR Report Key12159154
MDR Text Key262489184
Report Number9610612-2021-00503
Device Sequence Number1
Product Code FZP
UDI-Device Identifier04046963356040
UDI-Public4046963356040
Combination Product (y/n)N
PMA/PMN Number
K081031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 10/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/13/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberPL579T
Device Catalogue NumberPL579T
Device Lot Number52657392
Was Device Available for Evaluation? No
Date Manufacturer Received09/15/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-