The field service engineer (fse) conducted a site visit and was unable to confirm the reported issue by observing the bf wash probe 3.The fse was unable to reproduce the reported issue due to the visible damage to the probe.The fse resolved the issue by replacing the bf wash probe 3.The fse validated the instrument by performing quality control (qc).The qc results passed and was within published ranges.The instrument was returned to operational status.The aia-2000 analyzer is functioning as expected.No further action required by field service.A complaint and service history review for serial number (b)(4) from installation date of 14may2020 through aware date of 14jun2021 was performed for similar complaints.There were no similar complaints identified during the search period.The most probable cause of the reported event was failure of wash probe or ptfe tip.Submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.
|