• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 SYRINGE IRRIGATION BULK / CLEAN, 50ML; IRRIGATION SYRINGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 SYRINGE IRRIGATION BULK / CLEAN, 50ML; IRRIGATION SYRINGE Back to Search Results
Model Number 0935280
Device Problem Moisture Damage (1405)
Patient Problem Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 06/21/2021
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event Description
It was reported that there was moisture in the bulb syringe.
 
Manufacturer Narrative
The reported event was unconfirmed because the reported failure could not be reproduced.The reported failure is considered within specification as the reported failure could not be reproduced.The product was not used for patient treatment.The reported failure is considered within specification as the reported failure could not be reproduced.No root cause could be found because the reported event was unconfirmed.Visual evaluation of the returned sample noted one opened (without original packaging), 1 used irrigation syringe bulb.Visual inspection of the sample noted no moisture or damage found on the return sample.The device was returned for evaluation.A device history record review was not required as the investigation was unconfirmed.The investigation is concluded, and no additional action is required at this time.As the reported event is unconfirmed a risk review and labeling review is not required.The actual/suspected device was inspected.
 
Event Description
It was reported that there was moisture in the bulb syringe.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYRINGE IRRIGATION BULK / CLEAN, 50ML
Type of Device
IRRIGATION SYRINGE
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
yonic anderson
8195 industrial blvd
covington 30014
7707846100
MDR Report Key12174100
MDR Text Key261730787
Report Number1018233-2021-04235
Device Sequence Number1
Product Code KYZ
UDI-Device Identifier10801741051378
UDI-Public(01)10801741051378
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/15/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0935280
Device Catalogue Number0935280
Device Lot NumberNGEY0583
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/20/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/19/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/15/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-