BOSTON SCIENTIFIC CORPORATION INTELLANAV MIFI OPEN-IRRIGATED; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER
|
Back to Search Results |
|
Model Number 87047 |
Device Problem
Crack (1135)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/12/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device was returned to boston scientific for analysis.Visual inspection noted a cracked luer.Electrical tests revealed no electrical opens or shorts nd all electrodes, sensor and thermocouple resistances were in specification and were typical.All impedance and temperature readings were normal.The steering knob and the tension control knob functioned properly on both lock and unlock positions.No abnormal resistance was felt when actuating the steering mechanism.The device's lumen was leak tested using an isaac pressure decay test system set to 107 psi, and a touhy bourst seal for sealing the irrigation holes and me's.The lumen pressure decay was measured three times.Lumen pressure decay values indicated a gross leak due to the cracked luer.After the luer crack was sealed with adhesive lumen pressure decay values were below the maximum limit.X-ray examination revealed kink in the magnetic sensor wires located approximately 45.5 centimeters from the distal end.
|
|
Event Description
|
Reportable based on device analysis completed on june 16 2021.During an ablation procedure a intellanav mifi open-irrigated (oi) catheter was selected for use.It was reported that after approximately 45 minutes in vivo, high frequency and high amplitude noise was observed on electrogram channels on both rhythmia and the recording system.The noised ceased once the intellanav mifi oi catheter was disconnected from the new intellanav cable.The physician then reconnected the cable and the noise appeared again.There appeared to be some play in the connection and the physician felt a 'second click' when attaching the intellanav cable.The severity of the noise changed with manipulating the catheter handle at the junction of the port which connects the intellanav cable.The catheter was then replaced for another intellanav mifi oi catheter and the procedure was completed with no patient complications.However, device analysis revealed a cracked luer.
|
|
Search Alerts/Recalls
|
|
|