Model Number PL574T
Device Problems Fracture (1260); Separation Failure (2547)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/29/2021
Event Type  malfunction  
Manufacturer Narrative
Manufacturing site evaluation: investigation on-going.Additional information / investigation results will be provided in a supplemental report.
 
Event Description
It was reported that there was an issue with pl574t - challenger ti-p sm-ligat.Clips 12 cartr.According to the complaint description, the cartridge detached.The clip was not filled and the claw at the base of the cartridge broke and disappeared.The fragment remained in situ.Additional information was not provided.Additional patient information is not available.The malfunction is filed under aag reference (b)(4).Involved components: pl522r - shaft compl.D:5mm l:310mm - lot unknown.Pl520r - challenger ti-p handle - lot unknown.
 
Event Description
Involved components: pl522r - shaft compl.D:5mm l:310mm - lot unknown.Pl520r - challenger ti-p handle - lot unknown.
 
Manufacturer Narrative
Investigation results: visual investigation: the products arrived in a used condition.The investigation was carried out visually and microscopically.We made a visual inspection of the products, here we detected a deformed slider sheet and a broken off latch.In addition, we detected that the jaws are slightly bent and no longer according the dimensions according to our specification valid at the time of production (target: 3,3+0,1/ actual: 3,42 id: 33581).Batch history review: the device quality and manufacturing history records have been checked for the available lot number and were found to be according to our specifications valid at the time of production.Review of the complaint history revealed that no similar complaints have been filed against products from this batch number.The review of risk assessment revealed that the overall risk level (severity 4(5) x probability of occurrence 1(5)) according to din en iso (b)(4) is still acceptable.Conclusion and measures / preventive measures: based upon the investigation results a clear root cause conclusion cannot be drawn.There is no indication for a material-, manufacturing- or design-related failure.Based upon the investigations results a capa is not necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CHALLENGER TI-P SM-LIGAT.CLIPS 12 CARTR.
Type of Device
LIGATION / VESSEL CLIPS
Manufacturer (Section D)
AESCULAP AG
po box 40
tuttlingen, 78501
GM  78501
Manufacturer (Section G)
AESCULAP AG
po box 40
tuttlingen, 78501
GM   78501
Manufacturer Contact
christian von der grün
po box 40
tuttlingen, 78501
GM   78501
MDR Report Key12191592
MDR Text Key262487651
Report Number9610612-2021-00512
Device Sequence Number1
Product Code FZP
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K081031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative,Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/19/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberPL574T
Device Catalogue NumberPL574T
Device Lot Number52674383
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/14/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/14/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/01/2021
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
PL520R - LOT UNKNOWN; PL520R - LOT UNKNOWN; PL522R - LOT UNKNOWN; PL522R - LOT UNKNOWN
-
-