MAUDE Adverse Event Report: ST. JUDE MEDICAL, COSTA RICA LTDA TRIFECTA¿ GT VALVE; HEART-VALVE, NON-ALLOGRAFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, COSTA RICA LTDA TRIFECTA¿ GT VALVE; HEART-VALVE, NON-ALLOGRAFT TISSUE Back to Search Results
Model Number TFGT-23A
Device Problems Material Twisted/Bent (2981); Material Split, Cut or Torn (4008)
Patient Problem Dyspnea (1816)
Event Date 06/06/2021
Event Type  Injury  
Manufacturer Narrative
The results, method, and conclusion codes along with investigation results will be provided in the final report.
 
Event Description
It was reported that on (b)(6) 2018, a 23mm trifecta¿ gt valve was successfully implanted.On (b)(6) 2021, the patient presented with dyspnea.The device was explanted and replaced with a non-abbott device to resolve the event.Upon inspection of the explanted valve, 2 torn leaflets and a bent stent were noted.The physician thinks that the bent stent might've occurred during explant of the device.The patient is stable.No additional information was provided.
 
Manufacturer Narrative
Explant was reported due to dyspnea.The investigation found that leaflets 2 and 3 were torn.There was a thin lay of fibrin on all three leaflets.No acute inflammation, significant calcifications, or stent post deformation were present.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.This was inclusive of a review of the manufacturing videos, which contained no evidence of anomalies during functional inspection.In the absence of any calcification or evidence of infection, the reported event is consistent with a non-calcific leaflet tear.A non-calcific leaflet tear is a form of structural valve deterioration (svd), which is a well-known complication from valve replacement surgery.A non-calcific leaflet tear is commonly attributed to increased operational leaflet stress but may also be related to biological factors which result in tissue degeneration characterized by loss of collagen.In this case, histological evaluation did not demonstrate loss of collagen at the tear site and the cause of the leaflet tear could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIFECTA¿ GT VALVE
Type of Device
HEART-VALVE, NON-ALLOGRAFT TISSUE
Manufacturer (Section D)
ST. JUDE MEDICAL, COSTA RICA LTDA
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 01897 -405
CS  01897-4050
Manufacturer (Section G)
ST. JUDE MEDICAL, COSTA RICA LTDA
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 01897 -405
CS   01897-4050
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key12195957
MDR Text Key262423871
Report Number3014918977-2021-00027
Device Sequence Number1
Product Code LWR
UDI-Device Identifier05415067018229
UDI-Public05415067018229
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
P100029
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/20/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/06/2022
Device Model NumberTFGT-23A
Device Catalogue NumberTFGT-23A
Device Lot Number6752451
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/23/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received11/22/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/07/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;