It was reported that while performing an unspecified surgery with two units of floseal, the applicators reportedly broke and ¿lacerated the surgeon's hand¿.At the time of this report no further detail was provided regarding treatment, interventions for the event or the surgeon¿s outcome.No additional information is available.
|
Additional information: d9, h3, h6 and h10.Two devices were received for evaluation.It was reported that two applicators were broken; however, visual inspection was performed and no damage was observed on the applicators.It was identified that one side of the floseal syringe flange was broken off for both syringes.The location of the damage is not commonly observed.No additional issues were found during the sample analysis.One retention sample was reconstituted per the instructions for use (ifu).The thrombin solution was mixed into the gelatin syringe and the contents were mixed back and forth successfully between the syringes.The gelatin mixture was extruded out via the applicator with no issues were observed.The reported condition of broken applicator was not verified.The cause of the condition could not be determined; however, probable causes could be due to a user related handling issue.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|