Model Number M00510890 |
Device Problems
Separation Failure (2547); Material Twisted/Bent (2981)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/22/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The complainant was unable to provide the suspect device lot number; therefore, the lot expiration and device manufacture dates are unknown.(b)(4).The complainant indicated that the device was disposed and will not be returned for evaluation; therefore, a failure analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental mdr will be filed.
|
|
Event Description
|
It was reported to boston scientific corporation that a trapezoid rx lithotripter basket was used in the common bile duct during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2021.During the procedure, an alliance handle was used in conjunction with the trapezoid basket in an attempt to crush a stone.However, the handle cannula bent and the basket was unable to open.Additionally, the tip failed to separate from the basket to release the stone.With the stone still inside the basket, the physician pulled the scope and managed to pull the basket out.Procedure was completed.There were no patient complications as a result of this event.The patient's condition at the conclusion of the procedure was reported to be fine.
|
|