|
Model Number SGS-E2S |
Device Problems
Nonstandard Device (1420); Overheating of Device (1437)
|
Patient Problems
Burn(s) (1757); Erythema (1840)
|
Event Date 06/03/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
The dentist refused to provide the patient's weight.The same adverse event in this report has been reported to the fda separately by the initial importer, (b)(4), under report number (b)(4).
|
|
Event Description
|
On july 8, 2021, nakanishi became aware of a handpiece overheating through a complaint input into the complaint database by a distributor ((b)(4)).According to the distributor, there are two devices suspected to be involved in the event, but the dentist could not identify which one of the devices actually caused the following event.Therefore, nakanishi is submitting two separate mdrs for this event.This mdr is regarding the handpiece with the serial number (b)(4).- the event occurred on (b)(6) 2021.- the dentist was performing a tooth extraction on a patient using the sgs-e2s handpiece (serial no.(b)(4)).- during the procedure, the patient complained about heat on the patient's lower lip.- the dentist discontinued the procedure and felt the handpiece was very hot to the touch.- the dentist found some redness on the patient's lip with no blistering.- no medical treatment was required at the time of the event, and the patient has reported no complications.
|
|
Manufacturer Narrative
|
Upon receiving the device involved in the mdr event from the distributor, nakanishi conducted a failure analysis of the returned device, which included measuring the operating temperature of the device [report no.C210718-01].These activities are described in more detail below.Methodology used: a) nakanishi examined the device history record and the repair history for the subject sgs-e2s device [serial no.(b)(6)].There were no problems observed during manufacturing or testing noted in the dhr.There were also no repair history records since the device was shipped.B) nakanishi conducted temperature testing of the returned device in the following manner: b.1) temperature sensors were attached to the exterior of the device at various test points.This included the point most proximal to the patient (testing point (1)) and points further toward the distal end of the device (testing points (2) through (4)).The test setup was prepared to take temperature measurements at all points simultaneously, including a reference measurement at ambient room temperature.B.2) nakanishi attached a thermocouple (sensor to measure temperature) to each of the testing points.Nakanishi rotated the device's motor at 40,000 min-1, which is the maximum rpm for the motor that drives the handpiece (80,000 min-1 for the handpiece), with no water spray, and measured the exothermic response.B.3) nakanishi measured the temperature rise of the returned handpiece set at 80,000 min-1 (motor revolution 40,000 min-1).Nakanishi observed an abnormal temperature rise at test points (1) and (2) a few seconds into the test.Temperature measurements 50 seconds after the start of the test were as follows: - test point (1): 84.6 degrees c.- test point (2): 57.8 degrees c.- test point (3): 35.1 degrees c.- test point (4): 46.6 degrees c.The increase in temperature was so sudden that the test was concluded 50 seconds into the planned 5-minute evaluation period.Identification of the specific failure mode(s) and/or mechanism(s) of the associated device components was conducted as follows: a) nakanishi disassembled the handpiece and performed a visual inspection of the internal parts.Nakanishi observed the following: - the bearing retainer was broken.- the bearing was soiled and discolored.- there was a great amount of debris on the other internal parts.B) nakanishi took photographs of all the disassembled parts and kept them in investigation report no.(b)(4).Conclusions reached based on the investigation and analysis results: a) nakanishi determined that the cause of the overheating of the returned device was abnormal resistance during rotation due to the broken bearing retainer.Nakanishi considers the possibility from many years of experience that the cause of the broken bearing retainer was the ingress of undesirable materials into the bearing, leading to abrasion.B) a lack of maintenance caused the accumulation of debris on the internal parts, which caused debris ingress into the bearing during rotation.This contributed to the handpiece overheating.C) in order to prevent a recurrence of the handpiece overheating, nakanishi took the following actions: c.1) nakanishi reviewed the operation manual and reconfirmed the clarity and understandability of the instructions.C.2) nakanishi reported the above evaluation results to nsk america and directed nsk america to remind the user of the importance of maintenance, as instructed in the operation manual.
|
|
Search Alerts/Recalls
|
|
|