Model Number 10-80-00 |
Device Problem
Device Contamination with Body Fluid (2317)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/21/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Livanova (b)(4) manufactures the s5 roller pump.The incident occurred in (b)(6).Blood may have entered the pump, so the device will be replaced with an alternative one.Livanova initiated an investigation in order to determine if the reported event was related to the roller pump.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
Livanova (b)(4) received a report that during a procedure, a s5 roller pump 150 (usually used with a 10 mm tube) was used with a 6 mm suction tube which was cut and blood was scattered.The operation was completed without any problems.There was no report of patient injury.
|
|
Manufacturer Narrative
|
H.10: through follow-up communication with the customer livanova learned that the reported event could be traced to an user error.Reportedly, a 1/4 tube with a 3/8 tube insert was used and this caused the tube to get caught in the raceway and consequently burst.Based on the fact that the problem was solely caused by user error, the reported event has been reassessed as not reportable.
|
|
Event Description
|
See initial report.
|
|