• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND S5 ROLLER PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND S5 ROLLER PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 10-80-00
Device Problem Device Contamination with Body Fluid (2317)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/21/2021
Event Type  malfunction  
Manufacturer Narrative
Livanova (b)(4) manufactures the s5 roller pump.The incident occurred in (b)(6).Blood may have entered the pump, so the device will be replaced with an alternative one.Livanova initiated an investigation in order to determine if the reported event was related to the roller pump.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(4) received a report that during a procedure, a s5 roller pump 150 (usually used with a 10 mm tube) was used with a 6 mm suction tube which was cut and blood was scattered.The operation was completed without any problems.There was no report of patient injury.
 
Manufacturer Narrative
H.10: through follow-up communication with the customer livanova learned that the reported event could be traced to an user error.Reportedly, a 1/4 tube with a 3/8 tube insert was used and this caused the tube to get caught in the raceway and consequently burst.Based on the fact that the problem was solely caused by user error, the reported event has been reassessed as not reportable.
 
Event Description
See initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 ROLLER PUMP
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich
MDR Report Key12204039
MDR Text Key262693399
Report Number9611109-2021-00380
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 07/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/21/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10-80-00
Was the Report Sent to FDA? No
Date Manufacturer Received07/16/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-