• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SPINE INC CONFIDENCE SPINAL CMT SYS, 11C; POLYMETHYLMETHACRYLATE BONE CEMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SPINE INC CONFIDENCE SPINAL CMT SYS, 11C; POLYMETHYLMETHACRYLATE BONE CEMENT Back to Search Results
Model Number 283910000
Device Problem Migration or Expulsion of Device (1395)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/25/2021
Event Type  malfunction  
Manufacturer Narrative
Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Device report from synthes reports an event in (b)(6) as follows: it was reported that on (b)(6) 2021, the patient underwent for a vertebral arthrodesis procedure with pedicle screw cementation.During the surgery, the system that pushes the cement into the cannula / screw (pump system with water) did not work properly due to the water leaking.It was not possible to cement all the screws of the implant.The surgery was completed successfully with 30 minutes delay reported.There were no patient consequences are reported.Concomitant device reported: unknown screw (part# unknown, lot# unknown, quantity unknown).This report is for (1) confidence spinal cmt sys, 11c.This report is 3 of 3 (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: h3, h4, h6- the confidence spinal cmt sys, 11c (p/n: 283910000, lot #: unk) was returned and received at us cq.Upon visual inspection, only confidence pump assembly (p/n: 887014225) was returned from the kit.Some sterilized water was observed behind the piston assembly.The connector was inspected, and no issues were observed.The complaint condition was confirmed for the confidence spinal cmt sys, 11c (p/n: 283910000, lot #: unk) as no sterilized water was able to transfer through the connector because of the internal leakage into the pump body.There is no indication that a design or manufacturing issue has caused the issue and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: dhr cannot be performed as the lot number is unk.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONFIDENCE SPINAL CMT SYS, 11C
Type of Device
POLYMETHYLMETHACRYLATE BONE CEMENT
Manufacturer (Section D)
DEPUY SPINE INC
325 paramount drive
raynham MA 02767
MDR Report Key12206819
MDR Text Key262961975
Report Number1526439-2021-01460
Device Sequence Number1
Product Code NDN
UDI-Device Identifier10705034209623
UDI-Public10705034209623
Combination Product (y/n)N
PMA/PMN Number
K060300
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 06/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/21/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number283910000
Device Catalogue Number283910000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/17/2021
Date Manufacturer Received08/06/2021
Patient Sequence Number1
-
-