Device evaluation of electrode belt sn (b)(4) has been completed.The reported problem (gong alerts and wires exposed) has been confirmed.Upon investigation the electrode belt was unable to complete essential performace testing.The cable connecting the trunk cable to the distribution node (dn) was pulled from the strain relief, damaging trunk cable connector j702 on the distribution node pca.The root cause for the strained cable was excessive force.No adverse events occurred from the damaged electrode belt.
|