It was reported that during a security check at the airport, this pacemaker had a change in the programming of the sensors.Physician noted that the sensors were inactivated and proceeded to reactivate them.Everything is back to normal.The pacemaker remains in service.No additional adverse patient effects were reported.Boston scientific technical services (ts) analyzed the data and concluded that a handheld metal detector wand is unlikely to reprogram the device, it is possible for emi from a wand to induce noise resulting in sensor deactivation by the signal artifact monitor (sam).
|