Catalog Number 8065750597 |
Device Problem
No Display/Image (1183)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/30/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
|
Event Description
|
A customer reported that before the retinal surgery an ophthalmic laser presented with display issues.Surgery was completed using alternate with no reports of patient impact.
|
|
Manufacturer Narrative
|
The company service representative examined the system and replicated the reported event.The company service representative found the flex cable to cause a flickering issue.The nonconforming flex cable was replaced to resolve the issue.The system was then tested and met all product specifications.The system manufacturing device history record (dhr) was reviewed.Based on qa assessment, the product met specifications at the time of release.The root cause of the reported event can be attributed to a nonconforming flex cable.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
|
|