Model Number 5392 |
Device Problem
Failure to Calibrate (2440)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/23/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Product analysis: product analysis: analysis of the external pulse generator (epg) was unable to confirm the epg failed calibration.Analysis also noted that the epg powered up to an error code, the printed circuit board (pcb) was reset, the hanger assembly was broken, the device failed ventricular output dial at incoming functional testing, the encode assembly was replaced and the lower case was broken.All found defective parts were replaced and all other identified issues were resolved.The device passed all final functional tests.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the external pulse generator (epg) failed calibration.The peg was returned for service.There was no patient involvement.
|
|
Manufacturer Narrative
|
Following further analysis, it was found there was not enough information to make any determination based on the actual device.Review of the analysis report confirms encoder assy out of specification.The most likely root cause was traced to a component failure.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|