• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN EVERFLEX SELF-EXPANDING PERIPHERAL STENT WITH ENTRUST DELIVERY SYSTEM LONG; STENT, SUPERFICIAL FEMORAL ARTERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN EVERFLEX SELF-EXPANDING PERIPHERAL STENT WITH ENTRUST DELIVERY SYSTEM LONG; STENT, SUPERFICIAL FEMORAL ARTERY Back to Search Results
Catalog Number EVX35-08-120-150
Device Problems Inaccurate Delivery (2339); Activation, Positioning or Separation Problem (2906); Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/06/2021
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Physician was attempting to use the everflex entrust self-expanding stent along with a non medtronic 6fr sheath and non medtronic guidewire during procedure to treat soft tissue in the right common iliac artery at the proximal, mid and distal location.Tortuosity was severe and there was total chronic occlusion at 100%.The artery diameter was 8 mm - 10mm and lesion length was 20cm.The lesion was pre-dilated.There was no damage noticed to the packaging and no issues when removing the device from the hoop/tray.Ifu was followed.The everflex stent was implanted through the brachial artery and during stent implantation 12cm was extended to about 20cm.It was reported that stent deformation occurred in vivo during positioning and deployment.The catheter bunched up and excessive force was used during delivery of the stent.Stent struts were exposed during deformation.The stent was implanted in the patient.There was no vessel damage noted.The stent could not be implanted in intended location.The procedure was completed with 4 additional no n-medtronic devices.There was no patient injury reported.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVERFLEX SELF-EXPANDING PERIPHERAL STENT WITH ENTRUST DELIVERY SYSTEM LONG
Type of Device
STENT, SUPERFICIAL FEMORAL ARTERY
Manufacturer (Section D)
COVIDIEN
4600 nathan lane north
plymouth MN 55442
MDR Report Key12231194
MDR Text Key264289578
Report Number2183870-2021-00256
Device Sequence Number1
Product Code NIP
Combination Product (y/n)N
PMA/PMN Number
P110023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 08/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/27/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2022
Device Catalogue NumberEVX35-08-120-150
Device Lot NumberA808063
Was Device Available for Evaluation? No
Date Manufacturer Received08/02/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-