• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION AXIOS STENT AND DELIVERY SYSTEM; PANCREATIC STENT, COVERED, METALLIC, REMOVABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION AXIOS STENT AND DELIVERY SYSTEM; PANCREATIC STENT, COVERED, METALLIC, REMOVABLE Back to Search Results
Model Number M00553540
Device Problem Positioning Problem (3009)
Patient Problem Peritonitis (2252)
Event Date 06/18/2021
Event Type  Death  
Manufacturer Narrative
(b)(6).(b)(4).The complainant indicated that the device was disposed and will not be returned for evaluation; therefore, a failure analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental medwatch will be filed.
 
Event Description
It was reported to boston scientific corporation on (b)(6) 2021 that a hot axios stent was to be implanted transduodenal to the gallbladder during an endoscopic ultrasound (eus) guided gallbladder drainage procedure performed on (b)(6) 2021.During the procedure, the physician had difficulty deploying the stent because of low echographic visibility due to bleeding in the gallbladder which resulted in the stent being fully deployed in the duodenum.The stent was removed from the patient using forceps and another axios stent was implanted to complete the procedure.Later in the week after the hot axios implantation, the patient passed away due to a biliary leak.Reportedly, the patient was unable to undergo surgery due to other pathologies.In the physician's assessment, the axios stent deployed in an incorrect position did contribute to the biliary leak that led to patient's death.Additionally, the patient's comorbidity of severely extended cancer contributed to patient's death.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AXIOS STENT AND DELIVERY SYSTEM
Type of Device
PANCREATIC STENT, COVERED, METALLIC, REMOVABLE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key12254768
MDR Text Key264361391
Report Number3005099803-2021-03856
Device Sequence Number1
Product Code PCU
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
K150692
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 07/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/30/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/09/2023
Device Model NumberM00553540
Device Catalogue Number5354
Device Lot Number0026776814
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/09/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/09/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Death; Required Intervention;
-
-